SS-250
Ease of use and accuracy
Conveniently placed circular arm openings, a large front window and two side doors provide easy access for daily operations such as cleaning and disinfection. CO2 gas is supplied through the gas port located underneath the chamber, and gas concentration can be set from 0.0 to 10.0%

Features
Easy and safe handling of loaded ET catheters.
The large front door folds down for easy and safe handling of loaded ET catheters. Side doors on both sides allow for convenient passing in and out of materials. A rear door is also optional.

Clean and spacious fully flat work surface
The clean, spacious full-flat working surface is a single piece of heated stainless-steel with an integrated glass thermo plate, allowing for culture dishes to be slid across the surface without having to lift them, minimizing the risk of spillage. Integration of microscope optics and illumination can be made for most stereo microscopes available in the market.

Increase usable space without compromising your workspace.
A Utility Smart-Shelf on each side increases usable space without compromising the table top working space.


Specification
External Dimensions | 840 x 620 x 1190-1440 mm (W x D x H) |
Front Door Aperture | 710 x 190 mm (W x H) |
Side Door Aperture | 310 x 225 mm (W x H) |
Stage Height | 840〜1090 mm |
Weight | Approx. 80 kg (Scope excluded) |
Temp. Control Point | Working Stages: Right, Left, Center and Space |
Warm-up Time | 30 min. or more |
Wattage | 500W (120W at 37°C stable operation) |
Operating Environment Temp. | 20〜30°C |
Operating Humidity Limit | 20〜80%RH (non condensing) |
Power Source | AC110V, 115V, 220-240V 50/60Hz |
Working Stage: Right Left and Center
Temperature Control Range | Room temperature + 5°C -40°C |
Temperature Fluctuation Range | ±0.3°C (Ambient Temp.: 28°C, Set Temp.: 37°C) |
Temperature Distribution | ±0.5°C (Ambient Temp.: 28°C, Set Temp.: 37°C) |
Working Space
Working Space Dimensions | 826 x 420 mm (W x D) |
Temperature Control Range | Room temperature + 5°C - 40°C |
Temperature Fluctuation Range | ±0.5°C (Ambient Temp.: 28°C, Set Temp.: 37°C) |
Temperature Distribution | ±1.0°C (Ambient Temp.: 28°C, Set Temp.: 37°C) |
CO2 Control Range | 0.0 – 10.0% |
CO2 Concentration Fluctuation Range | ±0.2% |
CO2 Consumption | Approx. 170L/day (Approx. 7 days with 2.5kg cylinder) |
Air Control
Air Filter | 0.3um, Efficiency 99.97% or more |
Air Velocity | 0 – 0.5m/sec. variable |